Operating Manual Freeze Dryer BETA 1-8 LDplus / BETA 2-8 LDplus Maintenance and Service 61
7. Maintenance and Service
7.1. Maintenance
7.1.1. Ice Condenser Chamber
Before each start-up ensure that all water residue has been
removed from the ice condenser chamber. If necessary, wipe the
ice condenser chamber.
Before every drying process it is recommended to open and close
the drain valve once.
7.1.2. Heat Exchanger
A laminated heat exchanger is used to cool the refrigerant
compressed by the refrigerator. The heat exchanger is placed at
the back of the unit. In intervals of a few months it must be
checked for dust or dirt residues and cleaned if necessary.
Cleaning of the heat exchanger can be carried out best by
brushing, by using a vacuum cleaner from the outside or by using
compressed air from inside the unit.
Excessive dirt on the heat exchanger leads to a decrease in
performance and can lead to a failure of the unit!
Please note: The unit must not be turned upside down!
7.1.3. Vacuum Sensor VSP62/63
Please refer to the separate operating manual of the vacuum
sensor VSP62/63!
The VSP62/63 vacuum sensor is equipped with an internal Pirani
sensor, which utilises the thermal conductivity of gases. The
tungsten wire reacts with the components of the ambient gas,
which is why it only has a limited service life. Carbon-containing
substances, such as alcohol compounds in particular, extremely
reduce the service life.
The vacuum sensor is maintenance-free.
Remove any soiling on the outside with a cloth.
7.1.4. Rubber Valves
Special attention must be paid to the rubber valves. If the valves
are stiff, they must be dismantled, cleaned, slightly greased with
vacuum grease and reassembled.