32 Installation and Commissioning of the Unit Operating Manual Freeze Dryer BETA 1-8 LDplus / BETA 2-8 LDplus
4.4. Checking the Earth Connection
For checking the earth connection there is a screw for equalising
the ground potential on the back panel of the freeze-dryer. The
check can be carried out by means of an appropriate measuring
device.
4.5. Aeration
The aeration and drain valve at the lower left side of the unit is
used to vent the ice condenser chamber.
In case penicillin bottles have to be sealed under nitrogen instead
of vacuum, the ice condenser chamber can be flooded with
nitrogen via the nozzle of the aeration valve.
CAUTION! Max. 0.2 bar overpressure!
4.6. Condensate and Defrosting Water
The condensate and defrosting water is drained via the aeration
and drain valve on the lower left side of the unit. To drain the
water, connect the hose supplied with the unit to the nozzle. The
condensate and defrosting water are collected in a vessel.
The condensate and defrosting water can also be drained directly
via this hose. The condensate and defrosting water must be able
to drain freely. To achieve this, the hose must have a steady
downward slope. It must be ensured that water does not collect in
any part of the hose. The end of the hose always has to be above
the level of the liquid in the vessel for the discharged liquids.
Otherwise there is the risk of water and dirt residues being sucked
into the ice condenser chamber if there is a negative pressure
when opening the drain valve.
The drain valve is also used to vent the unit when a freeze-drying
process has been completed.
4.7. Vacuum Pump Exhaust
The oil mist which is produced during opferation of the vacuum
pump must be removed.
A ½" hose can be connected to the exhaust flange of the vacuum
pump RZ-2 or RC-5 and a ¾" hose is connected to the exhaust
flange of the vacuum pump DUO 5 or DUO 10. The hose either
leads into the open air or into a vent.
During installation of the pipe care must be taken that condensate
cannot flow back into the pump. With upward leading pipes it is
safest to use a separator (Woulfe's bottle or wash bottle) in the
pipe.
We strictly recommend using an exhaust filter (oil mist separator).
This filter prevents air pollution by oil mist which is emitted by the
vacuum pump in different quantities depending on the working
pressure.